The specific process of 80 VLF Hipot tester
Release Time : 2020-12-31 View Count : 次1) Self inspection process
The controller automatically enters the load detection. If no load is detected, the prompt message "load not connected" in the status bar of Figure 6 indicates that the booster or capacitive test sample is not connected.
Figure 6 the controller indicates that the load is not connected
1) Boosting process
After the self-test is successful, the instrument will automatically enter the boosting state, as shown in Figure 7, and the prompt message in the status bar is "boosting". At the same time, the timing began.
Figure 7 the controller indicates that the voltage is boosting
1) Equal amplitude output
The controller will raise the voltage to the set value in several cycles, and the instrument will output the equal amplitude, as shown in Figure 8. The prompt message in the status bar is "equal amplitude output".
Figure 8 constant amplitude output of controller prompt
4) Downtime
Figure 9 controller prompts to stop the test
When the timing reaches the set time, the instrument will stop automatically, as shown in Figure 9, and the prompt message in the status bar is "stop the test".
The instrument stops the high-voltage output and automatically discharges the test object, as shown in Figure 10. The prompt message in the status bar is "discharging".
Figure 10 the controller indicates that it is discharging
After the shutdown, as shown in Figure 11, the prompt message in the status bar is "test passed" and the data history is saved.
Note: during the test, if there is no abnormal voltage, no discharge or overcurrent protection, the test can be considered as passed.
Figure 11 controller prompt test passed
This instrument provides two shutdown modes:
★ timed shutdown: when the timing reaches the set time, the instrument will stop automatically.
★ manual shutdown: click the "shutdown" key to stop the machine. The two shutdown modes are normal shutdown.
★ there are also two kinds of abnormal shutdown: over-voltage protection shutdown and over-current protection shutdown.
★ overvoltage protection shutdown
In the process of test, when the output high pressure exceeds the set limit, the instrument will automatically cut off the output after starting and shutting down the command, and then execute the data history saving. After shutting down, as shown in Figure 12, the prompt message in the status bar is "over voltage protection" and execute the data history saving.
Fig. 12 controller prompts over voltage protection
★ over current protection shutdown
In the process of test, when the output current exceeds the set limit, the instrument will automatically cut off the output after starting and stopping the command, and then
execute the data history saving. After the shutdown, as shown in Figure 13, the prompt message in the status bar: "overcurrent protection" and execute the data history saving.
Fig. 13 over current protection prompted by controller
(1) printing
Click the "print" key in Figure 3 to print the data on the display into a test report. Under the status of viewing historical data, click the "print" key to print the historical data currently displayed on the screen.
(2) view historical data
Figure 14 historical data interface
All the data instruments that have been shut down by timing, clicking the "stop" key, over-voltage protection and over-current protection will automatically save them as historical data. It can save up to 64 times of measured data, and the previous 64 times will be deleted automatically. Click the "view" key in Figure 3 to open the interface in Figure 13 to view the historical data of the latest 64 tests.
(1) Clocksetting
Figure 15 date and time setting interface
Click the "clock" button in Figure 3 to open the setting interface in Figure 15, which is used to set the date and time of the system.